38
SimplyGo Mini User Manual
Visual, Audio
Indicators
Description What To Do
1 beep
that repeats every
16 seconds
Low Battery Alarm
This alarm occurs when
approximately 10 minutes
of battery life remains�
Remaining battery life is
dependent on your device
settings and your activity
level�
Press the yellow Alarm
silence button on the screen
to temporarily silence the
alarm, or press the Home
button on the device to
return to the Home screen,
and the 2nd screen will
appear with the Low Battery
symbol ashing on the
screen�
Note: This alarm could take
up to 60 seconds to activate
from the time the alarm
condition is present�
Replace the battery
or connect to a
power source�
Warm Up Indicator
This symbol indicates to
the user that the O
2
purity
has not yet been sampled
because the system needs
to stabilize� Its duration is at
least 14 minutes�
This symbol appears
onscreen after initial
startup and clears
from the screen
once the device has
reached the minimum
rated Oxygen
Purity� Contact your
equipment provider
if it persists on screen
for several hours after
startup�