LIFEPAK 500 automated external defibrillator Replacement Procedures
11-53
Installing Biphasic PCB—Biphasic AED
Application The following procedure describes the reinstallation of a Biphasic PCB that was
removed to gain access to other AED assemblies. For installation of a new
Biphasic PCB assembly, follow the Replacing Biphasic PCB—Biphasic AED
procedure.
To install the biphasic PCB:
1. Connect P106 of the Data Cable to J106 of the replacement Biphasic PCB.
2. Connect the seven spade lugs P101, P102, P103, P104, P105, P107, and
P108, to J101, J102, J103, J104, J105, J107, and J108, respectively, of the
replacement Biphasic PCB.
3. Reposition the Biphasic PCB over its well, gently moving and routing wires to
seat the Biphasic PCB without pressure against the Inductive Resistor coil or
the Biphasic PCB.
Jump to the Reseating Main PCB or Installing Main PCB—Biphasic AED
procedure.