LIFEPAK 500 automated external defibrillator Replacement Procedures
11-64
Replacing Speaker Assembly Page 3 of 3
Refer to the exploded view on the previous page.To install the new Speaker
Assembly:
1. Carefully clean the adhesive ring mating surfaces of the speaker cone, the
speaker assembly, and the inside of the case between the speaker assembly
tabs with a clean, lint-free cloth, dampened with isopropyl alcohol. Allow the
mating surfaces to dry thoroughly before going on to the next step.
2. Install the adhesive ring into the speaker cone.
3. Carefully rotate the replacement speaker and push it into the adhesive
ring to form a seal.
4. Dress the replacement speaker wires through the speaker cone wire
guide slot.
5. Check the speaker felts and, unless they are in good condition, slide two new
speaker felts into the case behind the sound ports.
6. Gently separate the walls of the assembly housing and slide the replacement
assembly into the lower enclosure. Make sure the flat side of the speaker
cone aligns with the bottom of the lower enclosure.
After installing the new Speaker Assembly, jump to the Reseating Main PCB
procedure and then to the Reassembling Case procedure.