LIFEPAK 500 automated external defibrillator Troubleshooting
8-6
Fault Code Table Page3of20
Code Fault Description
Corrective
Action
Code
5261 eeprom_tpu_channel_not_responding 2
5262 eat_eeprom_write_block_failure_1 2
5263 eat_eeprom_write_block_failure_2 2
5264 eeprom_hardware_problem_last 2
5265 eat_archive_qpdr_clash 6, 2
5266 eat_init_data_read_failure 6, 2
5267 eat_bad_papersize_crc 6, 2
5268 unreasonable_papersize_value 6
5269 eat_bad_cpss_crc 6, 2
526a eat_bad_asystole_time_crc 6, 2
526b unreasonable_manufact_config_analyze_key_value 6, 2
526c unreasonable_manufact_config_recording_equipped_value 6, 2
526d unreasonable_manufact_config_speaker_value 6, 2
526e eat-bad-manufact_config_key_crc 6, 2
526f unreasonable_energy_protocol_value 6, 2
5270 unreasonable_manufact_config_voltage_comp_value 6, 2
5271 unreasonable_energy_display_enabled_value 6, 2
5272 unreasonable_request_incident_id_enabled_value 6
Corrective Action Code Key
Code Description
1 Try to reload the software
2 Replace the Main PCB
3 Replace the Connector
4 Replace the Energy Storage Capacitor
5 Replace the Inductor
6 Recalibrate/Restore Setup, Complete PIP,
Return to service if fault codes do not recur
7 Check the Wires
8 Check the Battery
9 Replace the Time-of-Day Clock Battery
10 Clear Faults, Complete PIP, Return to service
if fault codes do not recur
11 Replace the Biphasic PCB