LIFEPAK 500 automated external defibrillator Troubleshooting
8-7
Fault Code Table Page4of20
Code Fault Description
Corrective
Action
Code
5273 eat_bad_last_incident_id_crc 6, 2
5274 eat_bad_alert_timer_option_crc 6, 2
5280 flash_fault_minor 10
5281 flashdrv_suspend_block_erase_wsm_busy 10, 2
5282 flashdrv_resume_block_erase_not_suspended 10, 2
5283 flashdrv_resume_block_erase_not_resumed 10, 2
5284 flashdrv_resume_block_erase_wsm_not_busy 10, 2
5285 flashio_cannot_clear_device 2
5286 flashio_erase_block_invalid_index 10, 2
5287 flashio_erase_complete_status_bad 10, 2
5288 flashio_erase_iterations_exceeded 10, 2
5289 flashio_initialize_unknown_device 2
528a flashio_write_byte_not_successful 2
528b flashio_write_byte_wsm_busy 2
528c flashio_wsm_not_ready 2
528d active_episode_info_duplicate_block_sequence_number 10
528e active_episode_info_block_sequence_number_out_of_range_1 10
528f active_episode_info_block_sequence_number_out_of_range_2 10
5290 active_episode_info_bad_block_detected_1 10, 2
Corrective Action Code Key
Code Description
1 Try to reload the software
2 Replace the Main PCB
3 Replace the Connector
4 Replace the Energy Storage Capacitor
5 Replace the Inductor
6 Recalibrate/Restore Setup, Complete PIP,
Return to service if fault codes do not recur
7 Check the Wires
8 Check the Battery
9 Replace the Time-of-Day Clock Battery
10 Clear Faults, Complete PIP, Return to service
if fault codes do not recur
11 Replace the Biphasic PCB