LIFEPAK 500 automated external defibrillator Troubleshooting
8-8
Fault Code Table Page5of20
Code Fault Description
Corrective
Action
Code
5291 event_retrieval_crc_fault 10, 2
5292 event_retrieval_incorrect_event_id 10, 2
5293 intermediate_results_retrieval_crc_fault 10, 2
5294 scene_audio_directory_entry_write_failure 10, 2
5295 scene_audio_block_state_write_failure 10, 2
5296 scene_audio_data_write_failure 10, 2
5297 scene_audio_block_state_status_write_failure 10, 2
5298 scene_audio_checksum_failure 10, 2
5299 obsolete_scene_audio_sample_number_out_of_order 10
529a summarization_ecg_decompression_error 10
529b summarization_z_decompression_error 10
529c summarization_ecg_compression_error 10
529d summarization_z_compression_error 10
529e summarization_ecg_wbtf_error 10
529f summarization_z_wbtf_error 10
52a0 summarization_wde_wbtf_error 10
52a1 waveform_wde_number_of_ecg_samples_mismatch 10
52a2 waveform_wde_number_of_z_samples_mismatch 10
Corrective Action Code Key
Code Description
1 Try to reload the software
2 Replace the Main PCB
3 Replace the Connector
4 Replace the Energy Storage Capacitor
5 Replace the Inductor
6 Recalibrate/Restore Setup, Complete PIP,
Return to service if fault codes do not recur
7 Check Wires
8 Check the Battery
9 Replace the Time-of-Day Clock Battery
10 Clear Faults, Complete PIP, Return to service
if fault codes do not recur
11 Replace the Biphasic PCB