LIFEPAK 500 automated external defibrillator Troubleshooting
8-10
Fault Code Table Page7of20
Code Fault Description
Corrective
Action
Code
5408 distrib_incorrect_size_raw_a2d_data 10
5409 distrib_incorrect_size_dsp_execution_count 10
540a distrib_incorrect_size_report_fault 10
540b dsp_comm_no_spectrum_data 10, 2
5500 rtc_fault_minor 9, 2
5501 rtc_autotest_not_scheduled 10, 9, 2
5502 rtc_nvram_crc_error 9, 2
5600 analyze_fault_minor 10
5601 mwb_samples_acquired_too_slow 10
5602 mwb_ecg_samples_out_of_order 10
5603 sas_task_dropped_leads_status_message 10
5604 sas_task_dropped_waveform_message 10
5605 sas_task_ecg_samples_not_contiguous 10
5606 incorrect_sizeof_waveform_data_type 10
5607 confirmation_of_ecg_samples_out_of_order_fault 10
5608 sas_task_voting_no_decision 10
5700 control_fault_minor 10
5701 control_missing_startup_switch_msg 10
Corrective Action Code Key
Code Description
1 Try to reload the software
2 Replace the Main PCB
3 Replace the Connector
4 Replace the Energy Storage Capacitor
5 Replace the Inductor
6 Recalibrate/Restore Setup, Complete PIP,
Return to service if fault codes do not recur
7 Check the Wires
8 Check the Battery
9 Replace the Time-of-Day Clock Battery
10 Clear Faults, Complete PIP, Return to service
if fault codes do not recur
11 Replace the Biphasic PCB