LIFEPAK 500 automated external defibrillator Troubleshooting
8-11
Fault Code Table Page8of20
Code Fault Description
Corrective
Action
Code
5702 control_unknown_dsp_msg_id 10
5703 control_start_inactivity_timer_failed 10
5704 control_unknown_interproc_data_category 10
5705 control_analysis_timer_expired 10
5706 control_dsp_shock_agree_timer_expired 6
5707 control_start_cpss_activation_timer_failed 10
5708 control_start_episode_commit_timer_failed 10
5709 control_start_prompt_display_timer_failed 10
570a control_start_clock_inactivity_timer_failed 10
570b control_start_cpr_timer_failed 10
570c control_start_shock_timer_failed 10
570d control_start_dsp_shock_agree_timer_failed 8, 6, 2
570e control_start_wait_for_charge_timer_failed 8, 6, 4, 2
570f control_start_analysis_timer_failed 10
5710 control_setup_first_shock_low_energy 6
5711 control_setup_first_shock_unknown_energy 6
5712 control_setup_second_shock_low_energy 6
5713 control_setup_second_shock_unknown_energy 6
Corrective Action Code Key
Code Description
1 Try to reload the software
2 Replace the Main PCB
3 Replace the Connector
4 Replace the Energy Storage Capacitor
5 Replace the Inductor
6 Recalibrate/Restore Setup, Complete PIP,
Return to service if fault codes do not recur
7 Check the Wires
8 Check the Battery
9 Replace the Time-of-Day Clock Battery
10 Clear Faults, Complete PIP, Return to service
if fault codes do not recur
11 Replace the Biphasic PCB