LIFEPAK 500 automated external defibrillator Troubleshooting
8-27
Troubleshooting Chart Page 4 of 8
Area Observation Suggested Corrective Action
Electrode Impedance Incorrect message response to
impedance settings
Verify test setup and retry test.
Check decade box operation and settings.
Check test cable continuity.
Perform TCP Battery Polarity Test.
Inspect patient/interface connector for
damage.
Check patient/interface cable connections
to Main PCB J2 and J4.
Replace Main PCB.
Energy Output and
Charge Time
Defibrillator will not charge Verify instructions and retry operation.
Verify a setup and retry test.
Use a fully functional battery.
Check connections between Energy
Storage Capacitor (C1- and C1+) and Main
PCB (P1 and P5).
Check or replace Energy Storage
Capacitor.
Replace Main PCB.