LIFEPAK 500 automated external defibrillator Troubleshooting
8-27
Troubleshooting Chart Page 4 of 8
Area Observation Suggested Corrective Action
Electrode Impedance Incorrect message response to
impedance settings
■ Verify test setup and retry test.
■ Check decade box operation and settings.
■ Check test cable continuity.
■ Perform TCP – Battery Polarity Test.
■ Inspect patient/interface connector for
damage.
■ Check patient/interface cable connections
to Main PCB J2 and J4.
■ Replace Main PCB.
Energy Output and
Charge Time
Defibrillator will not charge ■ Verify instructions and retry operation.
■ Verify a setup and retry test.
■ Use a fully functional battery.
■ Check connections between Energy
Storage Capacitor (C1- and C1+) and Main
PCB (P1 and P5).
■ Check or replace Energy Storage
Capacitor.
■ Replace Main PCB.