LIFEPAK 500 automated external defibrillator Troubleshooting
8-28
Troubleshooting Chart Page 5 of 8
Area Observation Suggested Corrective Action
Energy Output and
Charge Time (continued)
Defibrillator fails to discharge
Verify instructions and retry operation.
Verify test setup and retry test.
Check inductor connections to Energy
Storage Capacitor (P13) and transfer relay
(P14).
Check or replace Main PCB.
Delivered energy out of tolerance Verify instructions and retry operation.
Verify test setup and retry test.
Perform TCP Energy Calibration.
Check or replace Energy Storage Capacitor.
Check or replace Main PCB.
(LIFEPAK 500 Biphasic AED only): Replace
Biphasic PCB.
(LIFEPAK 500 Biphasic AED only): Replace
Inductive Resistor.