LIFEPAK 500 automated external defibrillator Troubleshooting
8-29
Troubleshooting Chart Page 6 of 8
Area Observation Suggested Corrective Action
Energy Output and
Charge Time (continued)
Charge to 360 J exceeds 15 seconds
Run first-Monday-of-the-month self-test.
Verify instructions and retry operation.
Use a fully functional battery.
Check or replace Battery Plate Wire
Harness.
Check or replace Energy Storage
Capacitor.
Check or replace Main PCB.
Automatic Advisory
Operation
Inappropriate or unintelligible screen
messages
Reload operating system (Physio-Control
only).
Check or replace Main PCB.
Inappropriate or unintelligible voice
prompts
Check or replace speaker.
Reload operating system (Physio-Control
only).
Check or replace Main PCB.
Improper response to analysis of
NSR input
Verify test setup and retry test.
Verify simulator delivers standard NSR
signal.
Contact Physio-Control Technical Services
for assistance.