LIFEPAK 500 automated external defibrillator Troubleshooting
8-30
Troubleshooting Chart Page 7 of 8
Area Observation Suggested Corrective Action
Automatic Advisory
Operation (continued)
Improper response to analysis
of VF input
Verify test setup and retry test.
Verify VF is present when analysis prompt
starts.
Verify simulator delivers standard VF
signal.
Contact Physio-Control Technical Services
for assistance.
Inappropriate second shock energy
level
Use Options Setup to establish desired
energy sequence.
Computer-Assisted Tests Failure to initialize automated test
application
Verify proper installation of computer-
assisted test routines.
UUT fails to respond to test stimuli
Verify instructions and retry operation.
Verify test setup and retry test.
Verify AED connection to an operational
computer COM port.
Use a fully functional battery.
UUT fails Automatic Tests Verify instructions and retry operation.
Check or replace Main PCB.