LIFEPAK 500 automated external defibrillator Troubleshooting
8-31
Troubleshooting Chart Page 8 of 8
Area Observation Suggested Corrective Action
Computer-Assisted Tests
(continued)
UUT fails Alarm Clock Test
Verify instructions and retry operation.
Replace Time-of-Day Clock battery.
Check or replace Main PCB.
UUT fails Audio Test (if equipped) Verify instructions and retry operation.
Check microphone.
Check or replace Main PCB.
UUT fails Power Failure Recovery
Test
Verify instructions and retry operation.
Check or replace Main PCB.
Leakage Current Excessive lead-to-lead leakage
current
Verify instructions and retry operation.
Verify test setup and retry test.
Check for and remove external leakage
current sources in the test setup vicinity.
Check, repair, or restore proper AED
internal wire routing.
Check or replace the Main PCB.