Communications Protocol CubiScan 125 Command Set
CubiScan 125 95
Scale Calibration Code Descriptions:
SA01 Clear the CubiScan
SA02 Place Test Weight in Corner
SA03 Scale Calibration Complete
Test
This command causes the CubiScan 125 to reply with an error code.
A response of TA00 means that the CubiScan 125 is ready and
responding to transmissions from the host. If the host receives no
response from the control unit after sending this command, an error
condition exists in the communications between the host and con-
troller.
Pos Len Description Type Range ASCII
Command Format
1 1 Start of Text Control (STX) 02h
2 1 Self Test Command Alpha (T) 54h
3 1 End of Text Control (ETX) 03h
4 1 Carriage Return Control (CR) 0Dh
5 1 Line Feed Control (LF) 0Ah
Acknowledge Format
1 1 Start of Text Control (STX) 02h
2 1 Self Test Command Alpha (T) 54h
3 1 Acknowledge Alpha (A) 41h
4 2 Identifier Numeric (00-99)
6 1 End of Text Control (ETX) 03h
7 1 Carriage Return Control (CR) 0Dh
8 1 Line Feed Control (LF) 0Ah

Table of Contents