Calibrating the Sensors Calibration
66 CubiScan 125
5. Place the calibration cube on the platform and slide it against
the back corner until it is in contact with both side panels with
the 12" side against the left side panel (when looking from the
front) as shown above.
6. Tap
[Next] to continue, and the following screen is displayed.
7. Turn the calibration cube so that the 12" side is against the right
side panel (make certain it is touching both side panels), as
shown above.
Figure 47
Third Sensor Calibration Screen