Communications Protocol CubiScan 125 Command Set
CubiScan 125 101
Zero
This command should be issued periodically to force the CubiScan
125 to perform internal compensations to adjust to changes in tem-
perature and humidity. This command should only be issued when
the measuring field is free of packages! Never issue this command
when a package is present in the measuring field.
Pos Len Description Type Range ASCII
Command Format
1 1 Start of Text Control (STX) 02h
2 1 Zero Command Alpha (Z) 5Ah
3 1 End of Text Control (ETX) 03h
4 1 Carriage Return Control (CR) 0Dh
5 1 Line Feed Control (LF) 0Ah
Acknowledge Format
1 1 Start of Text Control (STX) 02h
2 1 Zero Command Alpha (Z) 5Ah
3 1 Acknowledge Alpha (A) 41h
4 1 End of Text Control (ETX) 03h
5 1 Carriage Return Control (CR) 0Dh
6 1 Line Feed Control (LF) 0Ah
Negative Acknowledge Format
1 1 Start of Text Control (STX) 02h
2 1 Zero Command Alpha (Z) 5Ah
3 1 Neg. Acknowledge Alpha (N) 4Eh
4 1 End of Text Control (ETX) 03h
5 1 Carriage Return Control (CR) 0Dh
6 1 Line Feed Control (LF) 0Ah