Calibrating the Scale Calibration
68 CubiScan 125
IMPORTANT: Do not begin scale calibration until you have the test
weight. Calibrating without an accurate known weight (within .01
of a lb/kg) can make all future weight readings inaccurate.
Take the following steps to calibrate the CubiScan 125 scale.
NOTE
When calibrating the scale, the CubiScan 125 must be stable with
no movement of the platform such as that caused by vibration or
air movement.
1. Tap
[Menu] at the main screen.
☞
Figure 49
Main Screen