Calibrating the Scale Calibration
70 CubiScan 125
3. Make sure the CubiScan 125 platform is clear of all objects. Tap
[Scale]. The following screen is displayed.
4. If the weight units displayed are correct for the test weight you
are using, proceed to the next step.
If not, change the weight units as follows:
Tap the displayed weight unit. It will turn green. Then tap
[+] or
[–] to change the unit (e.g., from “lb” to “kg”). Tap the weight
unit again to save the change and return the display to normal.
5. If the calibration weight displayed matches the test weight you
are using, proceed to the next step.
If not, change the displayed weight as follows:
Tap the displayed weight until it turns green. Then tap
[+] or [–]
to change the weight one unit at a time. Tap the weight again to
save the change and return the display to normal.
Figure 52
First Scale Calibration Screen