Options in case when the main unit is [g]: g (gram), kg (kilogram), ct (carat),
lb (pound)*, oz (ounce)*, N (newton)*, u1 (custom unit 1)*, u2 (custom unit 2)*.
*) unit disabled for verified scales.
Options in case when the main unit is [kg]: kg (kilogram), lb (pound)*,
N (newton)*, u1 (custom unit 1)*, u2 (custom unit 2)*.
*) unit disabled for verified scales.
You can declare start unit and two custom units – for
detailed information read section 20, „UNITS”.
13.6. Setting MIN, MAX Thresholds
MIN, MAX thresholds are used:
to control mass of the weighed loads (read section 25.5 of this user
manual),
for graphic visualization informing you how much of the weighing device
capacity is used (read section 18.1.3 of this user manual),
to control the external automation systems using digital outputs of the
indicator (read section 17.2 of this user manual).
The mass is considered to be correct if it is comprised within MIN and MAX
threshold values. You can set MIN, MAX thresholds by:
using programmable <
Set MIN, MAX> button,
selecting product with set thresholds,
using a proximity sensor,
using a digital input.
MAX threshold value has to be greater than MIN threshold
value.
13.6.1. Setting MIN, MAX Threshold Using a Programmable Button
Enter <Display / Buttons functions> submenu.
Edit a given button.
Select <Set MIN and MAX> parameter from the list.
Exit to the home screen.
Press previously programmed button, <Min> numerical box is opened.