29.3. New Dosing Process
• Enter < Databases / Dosing processes> submenu.
• Press operation panel key to which
pictogram (add) is assigned. New
record is automatically edited.
• Name the dosing process, provide its code, go to a <Dosing process
step> submenu.
• Add the steps one by one in performance order, do it by pressing key
operating as
pictogram (add).
Dosing process steps:
Abbrev. Function Description
[DH] Dose manually
Enter to enforce manual weighing of the ingredient
(manual dosing).
Min Value of low threshold of manual dosing.
Max Value of max threshold of manual dosing.
Product Ingredient selected from the product database.
Confirm ingredients
manually
Mark to enforce manual confirmation of dosed mass by
means of pressing
key (the function is enabled for
manual dosing mode).
[DA] Dose automatically
Enter to enforce automatic weighing of the ingredient
(automatic dosing). The function allows to activate
outputs controlling dosing.
Fast dosing threshold
[DT1]
Value of mass for rough dosing in case of an automatic
2-stage dispensing.
Dosing threshold
[DT2]
Target mass value to be dosed in case of an automatic
dispensing.
Request mass
Query regarding value of mass that is to be dosed after
process start.
Product Ingredient selected from the product database.
[O] Outputs
Enter to set state of indicator outputs for activation
of external devices that are connected to these outputs.
Values: None – output inactive; „0” – LO state (output
off); „1” – HI state (output on).
[TI] Delay
Enter to determine time interval between successive
steps of the dosing process. The function defines queue
time for the next process step, given in seconds.
[Z] Zero
Enter to zero the platform, the function operates like
pressing of the
key placed on the indicator.
[T] Tare
Enter to tare the platform, the function operates like