Default code page value of the printer is 1250 – Central
European code page.
Exemplary weighing device settings for correct printout of Polish signs with use
of EPSON printer connected to RS232 port:
EPSON TM-U220D EPSON TM-T20 EPSON TM-T20
Baud rate
9600 bit/s 38400 bit/s 38400 bit/s
Parity
None None None
Code page
852 1250 852
Prefix
- 1B742D 1B7412
15.2.3. Prefix, Suffix
Prefix and suffix are controlling codes (given in a hexadecimal format) sent to
a printer at print start - <Prefix> parameter, and at print end - <Suffix>
parameter. Sending these codes allows to control globally information and
actions carried out at the beginning and/or at the end of each printout sent from
a scale to a printer. Most frequent usage:
Prefix - sending info on print code page.
Suffix - sending paper crop command in EPSON printers (in case
of printer equipped with an auto-cutter blade). Paper crop in EPSON
printer – code 1D564108.
Settings of <Prefix> and <Suffix> parameters are valid for all printouts sent via
the weighing device, e.g.: adjustment reports, statistics and header, footer and
GLP printouts.
15.2.4. Record of Measurement Data Onto a USB Flash Drive
Scale program enables record of measurement data onto an external flash
drive.
Procedure:
Plug a USB flash drive into the USB type A port.
Enter <Peripherals / Printer / Port> submenu and set <Pendrive> value.
Return to weighing.
Now the measurement data is saved to a text file on a USB flash drive
whenever you press
key. The file is generated automatically by
a scale software under printout.txt name.