27
Enter respective value and press key to confirm changes, <Max>
numerical box is opened.
Enter respective value and press
key to confirm changes.
13.6.2. Setting MIN, MAX Thresholds by Product Selection
Enter <Databases / Products> submenu.
Edit a given product and enter respective threshold values.
Exit to the home screen and press
button to select product with set
thresholds.
13.6.3. Setting MIN, MAX Threshold Using Proximity Sensor
Enter <Working modes> menu and select respective working mode.
Edit a given proximity sensor (left or right).
Select <Set MIN and MAX> parameter from the list.
Exit to the home screen.
Move your hand near the previously programmed proximity sensor, <Min>
numerical box is opened.
Enter respective value and press
key to confirm changes, <Max>
numerical box is opened.
Enter respective value and press
key to confirm changes.
13.6.4. Setting MIN, MAX Threshold Using a Digital Input
Enter <Inputs / Outputs> menu, next enter <Inputs> submenu.
Edit a given input.
Select <Set MIN and MAX> parameter from the list.
Exit to the home screen.
Activate the previously set input signal, <Min> numerical box is opened.
Enter respective value and press
key to confirm changes, <Max>
numerical box is opened.
Enter respective value and press
key to confirm changes.