73
Result control For detailed description read section 25.5.
Autotare For detailed description read section 25.6.
Labelling mode For detailed description read section 25.7.
Statistics For detailed description read section 25.8.
28.3. Reference Sample Mass Determined by Weighing
If the reference sample is to be weighed in a container, first put the container
on a weighing pan and tare it.
• Press
button (Determine reference sample mass), the following
message is displayed: <Put reference sample>.
• Load the weighing pan with the reference mass. When the indication is
stable (
pictogram is displayed) press key to confirm the
value.
• Now (instead of weight value of the measured load) difference between
values of loaded mass and reference mass is displayed in [%].
28.4. Reference Sample Mass Entered to the Scale Memory
• Press button (Set part mass), <Set reference sample mass> edit box
featuring an on-screen keyboard is displayed.
• Enter respective value and press
key to confirm.
• Now (instead of weight value of the measured load) difference between
values of loaded mass and reference mass is displayed in [%].
28.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part
from Database
Upon selecting a product from the database, reference sample mass of single
part assigned to the product (<Reference sample mass> entry) is entered
automatically.
Procedure:
• Enter <Percent weighing> mode and press operation panel key to which
pictogram (products database) is assigned, next select given product
from the list.
• Difference between values of loaded mass and reference mass assigned
to a selected product is displayed in [%].