22. ADJUSTMENT
Option available for non-verified scales exclusively
In order to ensure the highest weighing accuracy, it is recommended to
periodically introduce a corrective factor of indications to the weighing device
memory, the said factor must be referred to the reference mass. Adjustment
has to be carried out prior to the first weighing or if the ambient temperature
has changed dynamically. Prior to adjustment, unload the weighing pan.
22.1. External Adjustment
External adjustment is carried out using an external mass standard of the right
accuracy and weight value, which value depends on scale type and capacity.
Correction is carried out semi-automatically, successive process stages are
signalled with prompts.
Procedure:
• Enter <Adjustment / External adjustment> submenu. Message
<Remove weight> is displayed.
• Unload the weighing pan and press
key, message <Adjustment;
Please wait...> is displayed.
• Upon completed start mass determination, message <Put weight>
is displayed along with a particular value of mass standard assigned to the
scale.
• Load the weighing pan with the required adjustment weight and press
key.
• Upon completed procedure, message <Remove weight> is displayed.
• Take the weight off the weighing pan, the scale displays the
<Adjustment> submenu.
22.2. User Adjustment
User adjustment can be carried out using any adjustment weight, mass
of which ranges between: 0.3 Max - Max.
To run user adjustment procedure enter <Adjustment / User adjustment>
submenu. User adjustment and external adjustment procedures are likewise
with one exception, before user adjustment start, an edit box for entering mass
of the used adjustment weight is opened.
22.3. Start Mass Determination
If the scale does not require adjustment or you do not have suitable amount
of adjustment weights, you can determine start mass only.