69
ACAI For detailed description read section 27.2.1.
Minimum reference sample mass For detailed description read section 27.2.2.
Readout For detailed description read section 25.1.
Proximity sensors For detailed description read section 25.2.
Save mode For detailed description read section 25.3.
Auto threshold For detailed description read section 25.4.
Result control For detailed description read section 25.5.
Autotare For detailed description read section 25.6.
Labelling mode For detailed description read section 25.7.
Statistics For detailed description read section 25.8.
27.2.1. Function of Automatic Correction of Reference Sample Mass
<ACAI> is a specific function enabling you to correct weight value of a single
piece by means of the weighing device program.
Procedure:
• Enter <Working modes / Parts counting / ACAI> submenu and set
respective option.
Where:
Function of automatic correction of reference sample mass disabled.
Function of automatic correction of reference sample mass enabled.
<ACAI> function gets activated for <Parts counting> mode at the moment
of determination of reference sample quantity, it is signalled by display of
pictogram at the home screen top bar.
There are four conditions of ACAI function operation implemented into
the weighing device program:
1. the weighing result must be stable,
2. pieces quantity must increase,
3. pieces quantity upon adding cannot increase more than twice,
4. pieces quantity upon adding must be comprised within ± 0.3 tolerance of the
total value.