78
All process steps are performed, <Completed> status is displayed (dosing
process has been carried out). Blinking pictogram
is turned off.
29.4.1. Dosing Status
Dosing status provides information regarding the ongoing process step and is
comprised of 4 sections:
P1 1/10 DH -500g < [- - -]
1. Number of platform on which the process is carried out.
2. Information regarding the ongoing process step and quantity of all process
steps.
3. Abbreviation of process step type (for list of dosing process steps read
section 29.3), or global information regarding the process.
4. Detailed information regarding particular process step, where:
Step Information Description
[DH] -3.000kg < [Product 1]
-3.000kg - mass left to be dosed (in case of unstable mass
value at the step start, „?” status is returned). < - mass
status values: < - mass value lower than MIN threshold
value; > - mass value higher than MAX threshold value; OK
- mass between MIN, MAX thresholds pending the record;
RECORD – dosed mass record. [Product 1] - dosed
product name (lack of product is signalled by [---] status).
[DA] -3.000kg < [Product 1]
-3.000kg - mass left to be dosed (in case of unstable mass
value at the step start, „?” status is returned). < - mass
status values: < - mass value lower than the dosing
threshold value; OK - mass value higher than the dosing
threshold value, and pending the record; RECORD – dosed
mass record. [Product 1] - dosed product name (lack of
product is signalled by [---] status).
[O] 0111
Mask of set outputs: output 1 – high state, output 2 – high
state, output 3 – high state, output 4 – low state.
[TI] 00:00:05
Queue time for the next process step. Time given in format
HH:mm:ss (H – hour, m – minute, s – second).
[CM] Net >=1
Mass condition. The next step is to be performed when the
net mass value of the product on the platform is greater
than or equal to 1kg.