Remote Control
R&S
®
FPC
258User Manual 1178.4130.02 ─ 07
[SENSe:]CORRection:TRANsducer<td>:UNIT?
This command queries the unit of the current transducer factor.
Prerequisites for this command
Select a transducer ([SENSe:]CORRection:TRANsducer<td>:SELect).
Suffix:
<td>
.
Transducer
Return values:
<Unit>
Example: //Query the unit of the primary transducer
CORR:TRAN1:SEL 'HE200A-20-200.pritrd'
CORR:TRAN1:UNIT?
Usage: Query only
Manual operation: See "Transducer" on page 371
[SENSe:]CORRection:TRANsducer<td>[:STATe] <State>
This command turns a transducer factor on and off.
Prerequisites for this command
Select a transducer ([SENSe:]CORRection:TRANsducer<td>:SELect).
Suffix:
<td>
.
Transducer
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF | 0
Example: //Turn on primary transducer
CORR:TRAN1:SEL 'HE200A-20-200.pritrd'
CORR:TRAN1 ON
//Turn on secondary transducer
CORR:TRAN2:SEL 'PreAmp.sectrd'
CORR:TRAN2 ON
Manual operation: See "Transducer" on page 371
See "Transducer" on page 381
UNIT<n>:POWer <Unit>
This command selects the unit displayed on the y-axis.
Suffix:
<n>
.
irrelevant
Remote Commands of the Spectrum Application