CALCulate R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.52 E-6
CALCulate<Ch/Tr>:DATA:STIMulus?
Reads the stimulus values of the active data or memory trace.
<Ch/Tr>
Channel number used to identify the active trace.
Parameters
–
Range [def. unit] The data is transferred in the data format defined via
FORMat[:DATA]. The unit is the default unit of the sweep variable
(Hz or dBm or s).
*RST value –
SCPI,
Command
Types
Device-specific, query only
Example:
See CALCulate<Ch/Tr>:DATA?
CALCulate<Ch/Tr>:DLINe...
This subsystem controls the horizontal line used to mark and retrieve response values (display line).
CALCulate<Ch/Tr>:DLINe <numeric_value>
Defines the position (response value) of the horizontal line.
<Ch/Tr>
Channel number used to identify the active trace. If unspecified the
numeric suffix is set to 1.
Parameters
See list of parameters below.
Range [def.
unit]
The data is transferred in the data format defined via FORMat[:DATA].
The unit is the default unit of the measured parameter; see
CALCulate<Ch>:PARameter:SDEFine.
*RST value –
SCPI,
Command
Types
Device-specific, command or query.
Example:
*RST; CALC:DLIN 10
Define the position of the horizontal line in the default dB Mag diagram
at +10 dBm.
CALC:DLIN:STAT ON
Display the defined horizontal line.