Remote Control Commands
R&S
®
Scope Rider RTH
367User Manual 1326.1578.02 ─ 12
15.6.3.1 Configuring Harmonic Measurements
HARMonic:AVERage......................................................................................................367
HARMonic:CLEar...........................................................................................................367
HARMonic:DISPlay:TYPE...............................................................................................367
HARMonic:DISPlay:USER.............................................................................................. 368
HARMonic:FUNDamental:TYPE......................................................................................368
HARMonic:FUNDamental:USER..................................................................................... 369
HARMonic:FUNDamental:CURRent?...............................................................................369
HARMonic:LIMits:NAME.................................................................................................369
HARMonic:LIMits:LOAD..................................................................................................369
HARMonic:LIMits:CURRent?...........................................................................................370
HARMonic:LIMits:TYPE..................................................................................................370
HARMonic:SCALe..........................................................................................................370
HARMonic:STATistic.......................................................................................................370
HARMonic:THDType...................................................................................................... 371
HARMonic:AVERage <Average>
Determines the number of measurements for which the results are averaged. By
default, no averaging is performed.
Parameters:
<Average> OFF | AV2 | AV4 | AV8 | AV16 | AV32
*RST: OFF
HARMonic:CLEar
Clears the results for previous measurements used in statistical evaluation (see
HARMonic:AVERage on page 367 and HARMonic:STATistic on page 370).
Usage:
Event
HARMonic:DISPlay:TYPE <Selection>
Determines the number of harmonics to be displayed in the bargraph.
Note that this setting only affects the graphical result display; it has no effect on the
selection of harmonics for which results are provided during a file export (manually or
via remote command). For that purpose, use HARMonic:LIMits:TYPE on page 370.
Harmonics that were previously eliminated by the HARMonic:LIMits:TYPE com-
mand cannot be displayed.
Parameters:
<Selection> ALL | EVEN | ODD | ODD3 | ODN3 | USER
ODD3
All harmonics with an odd-numbered order that is a multiple of 3
ODN3
All harmonics with an odd-numbered order that cannot be divi-
ded by 3
Spectrum Analysis