1. Troubleshooting
8-7
Instrument condition
Probable cause
Action
Each door does not
move smoothly.
Foreign matter was entered into the
moving part of the door hinge, moving
part was deformed and damaged.
Check the operating part of the door hinge for pinching of foreign
matter or deformed/damaged hinge. If abnormality is found, remove
the foreign matter or replace the damaged part.
Deformed and damaged paraffin oven
door support.
If the door supports on the left and right sides of the paraffin oven
door are deformed, the door may contact the slits in the exterior
panels when opened/closed. If the deformation is significant,
remove the door and correct the deformation so that the slit will not
be contacted, or replace the door.
Door does not open.
Damaged catches of the fume control
unit door or reagent bottle rack door.
The catches on the fume control unit door and reagent bottle rack
door are of the push-open type. If these catches are damaged, the
door may not open even when pushed. Also, the door may not be
closed to the normal position. In this case, replace the damaged
catch.
Paraffin oven door getting stuck due to
paraffin.
If the paraffin oven door does not open, the door may have been
closed while dirty with paraffin and gotten stuck due to paraffin
deposits. Check to see if paraffin is deposited, and if paraffin is
deposited, remove it and then clean the surface with xylene, etc.
Entry of foreign matter
Foreign matter may be trapped somewhere under a moving part.
Check the instrument from the interior side and remove the foreign
matter.
Door does not close.
Inappropriately positioned or damaged
paraffin oven door magnet catch.
If the installation position of the magnet catch is changed or the
catch is deformed and damaged, the door may not close fully.
Check the condition of the magnet catch and make adjustment.
Replace the magnet catch if damaged, etc.
Detached paraffin oven door gasket.
The paraffin oven door gasket is of the insertion type. Accordingly,
opening the door when the gasket is stuck may cause the gasket to
come off and prevent the door from closing. In this case, clean the
gasket and insert it all the way into the groove.
Foreign matter on the catch receiving
part.
Foreign matter may attach to the catches on each door or the catch
contacting surfaces of the door and prevent the door from closing to
the normal position. In this case, remove the foreign matter.
Damaged catches of the fume control
unit door or reagent bottle rack door.
The catches on the fume control unit door and reagent bottle rack
door are of the push-open type. If these catches are damaged, the
door may not open even when pushed. Also, the door may not be
closed to the normal position. In this case, replace the damaged
catch.
External alarms are
not output.
(When the optional
"external interface
unit" is used.)
External alarms were not
communicated due to an external
wiring or connection equipment error.
Contact the Sakura Finetek Technical Support representative or
local representative.
External alarms are not output due to
an interface board relay error.
Contact the Sakura Finetek Technical Support representative or
local representative.
The connection harness is broken.
Contact the Sakura Finetek Technical Support representative or
local representative.
External alarms are not output due to
a CPU board error.
Contact the Sakura Finetek Technical Support representative or
local representative.
Rust generated.
Strong corrosive solution was used.
If very corrosive solution is used, the retort, etc., may rust. Use only
the permitted solutions. If you have any new solution you want to
use, contact the Sakura Finetek Technical Support representative
or local representative to see if it can be used.
Noise from the
instrument became
louder.
Rotating noise of the rotary valve.
Contact the Sakura Finetek Technical Support representative or
local representative.
Abnormal noise from the air pump.
Contact the Sakura Finetek Technical Support representative or
local representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Histo-Tek VP1 and is the answer not in the manual?

Sakura Histo-Tek VP1 Specifications

General IconGeneral
BrandSakura
ModelHisto-Tek VP1
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals