Cubis MSA User Manual 125
Calibration and Adjustment
y The procedure will be executed.
t Wait until calibration/adjustment has been completed.
y The results are displayed again after the procedure is completed.
t To return to operating mode, touch $CEM
Calibration/Adjustment Using External Check Weight
h
An external check weight is required for this function.
Please note the tolerance of the check weight being used.
t If required, switch to operating mode.
t Make sure that the weighing pan is empty.
J t Press 6#4', to tare the balance.
t Touch %#.
t Place the check weight on the balance.
t Touch 'ZVVGTPCNECNCFLYKVJHCEVQT[FGHKPGFYGKIJV.
y Calibration is carried out (approx. 15 to 20 sec.).
y After calibration, the determined deviation is displayed.
t To adjust the balance, touch #FLWUV
t To complete calibration without adjustment, touch 'ZKV.