Cubis MSA User Manual 149
Interfaces
Data Output
You can define the data output parameter so that output is activated either when
a print command is received or automatically synchronized with the display or at
defined intervals (see application programs and autoprint settings).
Data Output by Print Command The print command can be transmitted by pressing P or by a software
command (Esc P).
Automatic Data Output In the Autoprint mode, data is output to the data interface port without an extra
print command. You can have synchronized data output automatically at defined
display update intervals, with or without the stability parameter. The interval time
depends on the balance operating status and balance type.
If the automatic data output is activated in the Device Configuration, it starts
immediately after the balance is turned on. You can also configure whether the
automatic data output can be stopped and started by pressing the P key.
Data Output Formats
You can output the value displayed in the measured value line and the weight
unit, with or without a data ID code. Configure this output parameter in the
Device Configuration menu (Menu/Device Configuration/Configure data output/
Line format).
Example: Output Without an ID Code + 253 pcs 16 characters are output
Example: Output Without an ID Code Qnt + 253 pcs 22 characters are output
Data Output Format with 16 Characters
Display segments that are not activated are output as spaces.
Display values without a decimal point are output without a decimal point.
The type of character that can be output depends on the character’s position.
Normal Operation
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ A A A A A A A A A * E E E CR LF
or . . . . . . . . . . * * *
or * * * * * * * * * *
*: Spaces CR: Carriage return