EasyManuals Logo

Sartorius MSA3.6P User Manual

Sartorius MSA3.6P
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Cubis MSA User Manual 17
Getting Started
Installation Instructions
t Select the proper setup location:
Set up the device on a stable, even surface that is not exposed to vibrations
(e.g. weighing stone).
Place the device in a location with enough free space around it so that
excessive heat cannot build up.
Maintain free access to the device at all times.
Choose a location that is not subject to the following negative influences:
Heat (heater or direct sunlight)
Drafts from open windows, AC systems and doors
Extreme vibrations during weighing
Excessive moisture
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when it is
brought into a substantially warmer area.
To avoid the effects of condensation, condition the balance for about two hours at
room temperature, leaving it unplugged from the power supply.
Carefully read all warnings and safety precautions in the respective section of this
manual.
Setting the Angle for the Display and Control Unit
The angle of the display and control unit can be adjusted by the user, in order to
ensure optimal readability of the weight values on the display at all times.
t On MSA and MSU models, the display and control unit can be tilted as desired.
Power Cord Receptacle
t Check the voltage rating and plug design.
Make sure that the voltage rating printed on the manufacturer’s ID label is
identical to that of your local line voltage.
If the stated supply voltage or the plug design of the power cord does not
comply with your country’s standard, please inform the nearest Sartorius
representative or your dealer.
The power connection must be made in accordance with the regulations
applicable in your country.
In order to connect the equipment to the power supply (protection class 1),
use a correctly installed wall outlet with a protective grounding conductor (PE)
and a fuse of maximum 16 A.
If a main feeder from the ceiling is required or if installation of a CEE plug is
necessary, these must be carried out by a qualified electrician.
Only use original Sartorius AC adapters.
°C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MSA3.6P and is the answer not in the manual?

Sartorius MSA3.6P Specifications

General IconGeneral
BrandSartorius
ModelMSA3.6P
CategoryScales
LanguageEnglish

Related product manuals