154 Cubis MSA User Manual
Interfaces
Data Input
SBI Commands (Data Input Format)
The computer connected via the data port can send commands to the balance to
control balance and application program functions.
These control commands may have different formats and contain up to
26 characters. Each of these characters must be sent based on the setup
configuration for data transmission.
Formats for Control Commands
Format 1: Esc ! CR LF
Format 2: Esc ! # _ CR LF
Format 3: Esc ! # & (max. 20 &) & _ CR LF
Esc: Escape
!: Command character
#: Number
&: Number or letter
_: Underline (ASCII: 95)
CR: Carriage return (optional)
LF: Line feed (optional)
max: Depends on command character (parameters): The entry is truncated after
the max. length, and not rejected as when entered via the keyboard.
Examples:
Format 1: Esc P
Format 2: Esc x1_
Format 3: Esc t Article1_
Format 4: Esc e!{MESSAGE.SHOW_ERROR “Weight too low!}_
1
)
Esc e?{WEIGH.WGT_NET}_
1
)
(e.g. query net weight value)
Esc ea{Art.413}_
1
)
(determine weight and tare values and save to the alibi memory
together with the label)
1
) Additional parameters and module identifications are available in the document
“SBI Interface” as a download from the Sartorius homepage.