Cubis MSA User Manual 105
Timer-controlled Functions
Timer-controlled Functions
Purpose: This application is used to trigger individual balance functions automatically at
a specific time or after specific time intervals. The time or interval must be within
24 hours (1 day).
Options: You can select the following functions:
- Acoustic signal
- Lock in readout
- Automatical printout of the display value
- Transfer of components for totalizing, formulation or statistics
Example: In order to determine the amount of evaporation of a sample over time, you can
place the sample on the weighing pan and print out the weight value at fixed
intervals (e. g. every two minutes).
Combination options: Totalizing, statistics, formulation, mass unit conversion, second tare, memory,
individual identifiers (SQmin function, DKD measurement uncertainty)
Configuring Task: Timer-controlled Functions
D
t Define which function should be timer controlled.
#EQWUVKEUKIPCNDGGRVQPG The balance makes a beeping sound.
.QEMKPTGCFQWV The measured weight value
remains on the display for the defined time period.
#WVQOCVKERTKPVQWVQHXCNWGU: The displayed value is printed at the
defined time.
t Select the mode.
6KOGTUGVVQKPVGTXCNUThe function is triggered at regular intervals
(e.g. every 30 sec.).
6KOGTUGVVQCURGEKHKEVKOG The function is triggered at the fixed time
(after the start of the application) (e.g. at 08:00:00).
t Specify when the time and/or interval should be entered.
&GHKPGHKZGFVKOGYou must enter this timepoint and/or interval now.
6KOGUGVVKPIECPDGEJCPIGF: The time and/or interval is only entered when
the task is carried out.