124 Cubis MSA User Manual
Calibration and Adjustment
Calibration and Adjustment
Background During calibration, a check weight is used to determine how much the displayed
value deviates from the actual measurement value. This deviation is compared
with an entered target value and this deviation is then eliminated by a subsequent
adjustment of the balance. During the Linearization, the deviation of the values
from the ideal characteristic curve is cleared.
Not all functions/settings can be selected on balances used as legal measuring
instruments!
When and how often Calibration/Adjustment should be carried out on a regular basis, e. g. daily after the
balance is turned on. In addition, it should be done after each leveling and always
when the ambient conditions have changed (temperature, humidity or air pressure),
and when the balance has been set up in another location.
isoCal All Cubis balances are equipped with the automatic calibration function isoCal
(time and temperature-controlled). You can adapt this function according to your
requirements. You can set up a memory function that triggers a prompt for
calibration/adjustment at specified intervals.
h
The process for the calibration/adjustment function may differ depending on the
presets (see System Settings/Configuring Calibration/Adjustment). The factory
settings have not been changed for the procedure described here.
Calibration/Adjustment Using Internal Check Weight
t If required, switch to operating mode.
t Make sure that the weighing pan is empty.
J t Press 6#4', to tare the balance.
t Touch %#.
t Touch +PVGTPCNECNCFLWUVOGPV and then 5VCTV.