178 Cubis MSA User Manual
Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Semi-microbalances 0.01 mg
Model MSA225S-0CE MSA225P-0CE MSA125P-0CE
Accuracy class* k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 0.01 0.01/0.02/0.05 0.01/0.1
Weighing capacity max* g 220 60/120/220 60/120
Calibration value e* mg 1 1 1
Min. load min* mg 1 1 1
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.001 – 220 0.001 – 220 0.001 – 120
Min. initial weight** mg 20 20 20
Typical stabilization time s 2 2 2
Typical measurement time s 6 6 6
External standard calibration value g 200 (E2) 200 (E2) 100 (E2)
(min. accuracy class)
Application range (temperature) With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +15 to +25 °C
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display result (depending on the set filter level) 0.2 – 0.4
Weighing pan size (W D) mm 85 85
Weighing chamber height (draft shield DU)
mm 261
Protection Protected against dust and water
Verified models with EC Type Approval Certificate: Analytical balances 0.1 mg
Model MSA524S- MSA524P- MSA324S- MSA224S- MSA324P- MSA124S-
0CE 0CE 0CE 0CE 0CE 0CE
Accuracy class* k k k k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 0.1 0.1/0.2/0.5 0.1 0.1 0.1/0.2/0.5 0,1
Weighing capacity max* g 520 120/240/520 320 220 80/160/320 120
Calibration value e* mg 1 1 1 1 1 1
Min. load min* mg 10 10 10 10 10 10
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.01–520 0.01–520 0.01–320 0.01–220 0.01–320 0.01–120
Min. initial weight** mg 120 120 120 120 120 120
Typical stabilization time s 1 1 1 1 1 1
Typical measurement time s 3 3 3 3 3 3
External standard calibration value g 500 500 200+100 200 200 +100 100
(min. accuracy class) (E2 (E2) (E2) (E2) (E2) (E2
Application range (temperature) With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +15 to +25 °C
Display result (depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 85 85
Weighing chamber height (draft shield DU)
mm 261
Protection Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26