Cubis MSA User Manual 109
DKD Measurement Uncertainty
DKD Measurement Uncertainty
Purpose: This application is used to ensure that the measurement uncertainty is displayed
dynamically so that it conforms to the data documented on the DKD calibration
certificate.
Requirements: The function can only be used when the balance has been prepared for this before-
hand by a service technician. The service technician determines the measurement
uncertainty of the balance via a DKD calibration onsite. The technician records on
the DKD calibration certificate the measurements and the measurement uncertainty
for initial weighing. He also saves the calculated data on the balance.
Options: The measurement uncertainty can be displayed as an absolute value (U), a relative
value (U*) or as process accuracy (PA).
Combination options: Mass unit conversion, second tare memory, individual identifiers, (SQmin function)
Configure Task: DKD Measurement Uncertainty
D
t Select the measurement uncertainty display mode in relation to the maximum
capacity.
#DUQNWVGWPEGTVCKPV[: Absolute value (e. g. 12.00 g)
4GNCVKXGWPEGTVCKPV[ Relative value (e. g. 0.000045%)
2TQEGUUCEEWTCE[ e.g. 0.00013%
t If you have selected process accuracy, now enter the process accuracy factor.