Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances
Models MSA5203S-0CE MSA5203P-0CE MSA3203S-0CE MSA3203P-0CE
Accuracy class* k k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 1 1/2/5 1 1/10
Weighing capacity max* g 5200 1200/2400/5200 3200 1010/3200
Calibration value e* mg 10 10 10 10
Min. load min* mg 100 100 100 100
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.1 – 5200 0.1 – 5200 0.1 – 3200 0.1 – 3200
Min. initial weight** g 1.5 1.5 1.5 1.5
Typical stabilization time s 1 1 1 1
Typical measurement time s 2 2 2 1.5
External standard calibration value g 5000 (E2 5000 (E2 2000 (E2 2000 (E2)
(min. accuracy class)
Application range (temperature) With “isoCAL” function: +5 ... +40 °C
Without “isoCAL” function: +15 ... +25 °C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 140 140
Weighing chamber height (draft shield DE)
mm 172
Protection Protected against dust and water
Models MSA2203S-0CE MSA2203P-0CE MSA1203S-0CE
Accuracy class* k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 1 1/10 1
Weighing capacity max* g 2200 1010/2200 1200
Calibration value e* mg 10 10 10
Min. load min* mg 100 100 100
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.1 – 2200 0.1 – 2200 0.1 – 1200
Min. initial weight** g 1.5 1.5 1.5
Typical stabilization time s 1 1 1
Typical measurement time s 1.5 1.5 1.5
External standard calibration value g 2000 (E2) 1000 (E2) 1000 (E2)
(min. accuracy class)
Application range (temperature) With “isoCAL” function: +5 ... +40 °C
Without “isoCAL” function: +15 ... +25 °C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 140 140
Weighing chamber height (draft shield DE)
mm 172
Protection Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26

Table of Contents