EasyManua.ls Logo

Sartorius MSA3.6P - Special Outputs; Error Message

Sartorius MSA3.6P
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
150 Cubis MSA User Manual
Interfaces
Special Outputs
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * * * * – * * * * * * CR LF
or* * * * * * A * * * * * * * CR LF
or* * * * * * A B * * * * * * CR LF
and only upon request with ESC w0 (no print command):
* * * * * W * X X X Y Y Y CR LF
or* * * * * * I * X X X * * * CR LF
*: Spaces A = C: Adjustment
AB = – –: Final readout W: Draft shield status (optional):
A = H: Overload I: Ionizer (optional)
AB = H H: Overload in checkweighing Y,Y,Y = Draft shield doors
A = L: Underweight XXX = Decimal value calculated from
AB = L L: Underweight in checkweighing binary data
Decimal value Binary value Control information
1
Bit0 = 0: No error/ionizer off
Bit0 = 1: Draft shield error/ionizer on
2
Bit1 = 0: Draft shield motors off
Bit1 = 1: Draft shield in motion
8
Bit3 = 0: Learning function off
Bit3 = 1: Learning function on
16
Bit4 = 0: At least one draft shield door open
Bit4 = 1: All draft shield doors closed
32
Bit6 = 0: Motorized draft shield operation
Bit6 = 1: Manual draft shield operation
R,M,L = COO: Right door closed(Closed), Middle and Left doors open (Open)
R,M,L = OCC: Right door open (Open), Mmiddle and Left doors closed (Closed)
Error Message
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * E r r * */# # # * * * * CR LF
*: Spaces # # #: Error code number
Example: Output of the weight value + 111.255 g
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * 1 1 1 . 2 5 5 * g * * CR LF
Position 1: Plus +, or minus – or space
Position 2: Space or weight value
Position 3 – 10: Weight value with decimal point, leading zeros are output as
spaces.
Position 11: Spaces
Position 12 – 14: Characters for unit of measure or space
Position 15: Carriage return
Position 16: Line feed

Table of Contents