Specifications
Model-specific Data
Semi-microbalances 0.01 mg
Model MSA225S MSA225P MSA125P
Readability mg 0.01 0.01/0.02/0.05 0.01/0.1
Weighing capacity g 220 60/120/220 60/120
Tare range (subtractive) g – 220 – 220 – 120
Repeatability ±mg 0 to 6 g: 0.015 0 to 6 g: 0.015 0 to 6 g: 0.015
60 to 220g: 0.025 60 to 220g: 0.04 60 to 120g: 0.06
Linearity ±mg 0.1 0.15 0.15
Corner load (test load [g]) mg 0.15 (100) 0.2 (100) 0.15 (50)
Min. initial weight* mg 20 20 20
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1
Typical stabilization time s 2 2 2
Typical measurement time s 6 6 6
External standard calibration value g 200 (E2) 200 (E2) 100 (E2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.2 – 0.4
Weighing pan size (W D) mm 85 85
Weighing chamber height (draft shield DU)
mm 261
Protection Protected against dust and water
* = Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26
Analytical balances 0.1 mg
Model MSA524S MSA524P MSA324S MSA224S MSA324P MSA124S
Readability mg 0.1 0.1/0.2/0.5 0.1 0.1 0.1/0.2/0.5 0.1
Weighing capacity g 520 120/240/520 320 220 80/160/320 120
Tare range (subtractive) g – 520 – 520 – 320 – 220 – 320 – 120
Repeatability <±mg 0.1 0.15/0.2/0.4 0.1 0.07 0.1/0.2/0.4 0.1
Linearity <±mg 0.4 0.5 0.3 0.2 0.5 0.2
Corner load (test load [g]) mg 0.3 (200) 0.4 (200) 0.3 (200) 0.2 (100) 0.4 (200) 0.2 (50)
Min. initial weight* mg 120 120 120 120 120 120
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1 1 1 1
Typical stabilization time s < 1 < 1 < 1 < 1 < 1 < 1
Typical measurement time s < 3 < 3 < 3 < 3 < 3 < 3
External standard calibration value g 500 500 200+100 200 (E2) 200+100 100 (E2)
(min. accuracy class) (E2) (E2)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 85 85
Weighing chamber height (draft shield DU)
mm 261
Protection Protected against dust and water
* = Position according to OIML R76

Table of Contents