176 Cubis MSA User Manual
Specifications
Model-specific Data
Precision balances
Models MSA6202S MSA6202P MSA5202S MSA4202S
Readability mg 10 10/20/50 10 10
Weighing capacity g 6200 1500/3000/ 5200 4200
6200
Tare range (subtractive) g – 6200 – 6200 - 5200 – 4200
Repeatability ±mg 7 7/20/40 6 7
Linearity ±mg 20 50 10 20
Corner load (test load [g]) mg 20 (2000) 50 (2000) 10 (2000) 30 (2000)
Min. initial weight* g 12 12 10 12
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 2 2
Typical stabilization time s 1 1 0.8 0.8
Typical measurement time s  1.5 1.5 1 1
External standard calibration value kg 5 (E2) 5 (F1) 5 2 (E2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 206 206 206 206 140 140 206 206
Protection Protected against dust and water
* = Position according to OIML R76
Models MSA2202S MSA1202S MSA12201S MSA8201S MSA5201S
Readability mg 10 10 100 100 100
Weighing capacity g 2200 1200 12,200 8200 5200
Tare range (subtractive) g – 2200 – 1200 – 12,200 – 8200 – 5200
Repeatability ±mg 7 7 50 50 50
Linearity ±mg 20 20 100 100 100
Corner load (test load [g]) mg 20 (1000) 20 (500) 200 (5000) 200 (5000) 200 (2000)
Min. initial weight* g 12 12 100 100 100
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 4 4 4
Typical stabilization time s 0.8 0.8 0.8 0.8 0.8
Typical measurement time s 1 1 1 1 1
External standard calibration value kg 2 (F1) 1 (F1) 10 (F1) 5 (F2) 5 (E2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 206 206
Protection Protected against dust and water
* = Position according to OIML R76