Cubis MSA User Manual 19
Getting Started
Leveling the Balance, Setting the Level Indicator
h
Leveling the balance compensates for slant or unevenness at the place
of installation. The balance must be perfectly horizontal to ensure
consistent, reproducible weighing results. All models are equipped with
an electronic tilt angle detection feature. If the balance is not level, all
balance models issue an alert on the display.
Push-button automatic leveling is available on models with motorized
leveling feet. For models with manual leveling feet, please follow the
instructions on the display.
Leveling the Balance Manually
t The balance is leveled using both front leveling feet.
t Screw in both back leveling feet (only for models with back leveling feet).
t Turn the two front feet as shown in the illustration until the air bubble is
centered within the circle of the level indicator.
y Normally, several leveling steps are required.
t Screw out both back leveling feet until they touch the setup surface
(only for models with back leveling feet).
Warm-up Time
y To deliver precise results, the balance must warm up for at least 30 minutes
after initial connection to the power supply. Only after this time will the device
have reached the required operating temperature.
When a verified balance used in legal metrology (legal-for-trade
applications) is connected to the power, it must warm up for at least
two hours before operation.
t Turn on devices via the A key. The following appears in the display:
y You can now follow the brief instructions to configure balance settings before
beginning weighing operations.