Cubis MSA User Manual 27
Getting Started
2. Push the cover of the below-cell weighing port to one side.
3. Carefully screw on the below-cell weighing hook.
3
Do not screw it in too tightly,
as this could damage the thread or the balance.
h
Install a draft protection shield.
t Attach the sample (e.g., using a suspension wire) to the hook.
4 When weighing has been completed, unscrew the hook and return it to the clip.
t Close the cover of the below-cell weighing port.
The below-cell weighing port may not be opened or used on balances
used for legal metrology.
Balances with a readability of ) 1 μg
t Remove both screws from beneath the weigh cell and detach the cover plate.
t Attach a wire, for example, to the sample and hang it on the notched hook.
h
Install a draft protection shield.
2
3
4