Cubis MSA User Manual 89
Calculation
Executing a Task: Calculation “Air buoyancy correction”
t If you haven’t already done so, go to Task Management.
y The task selection is displayed.
t Touch the „Air buoyancy correction“ task from the Task menu.
y The formula for the air buoyancy correction is already defined.
t Enter the density of the sample via „Variable X,“ in this example 2.400 g/cm
3
.
t Touch 5VCTV to begin the calculation.
t The calculation results are displayed.
t To return to the weight value display, touch 9GKIJ.
t You can view both values, the weight and the mass.
t If you want to carry out additional weighing and calculations using
the same formula, place the sample on the balance and then touch %CNE.
t If you want to carry out additional weighing using a different formula,
touch $CEM.
t You can now change the formula and then proceed as described above.
Print via the P key:
N + 3.183629 g
X 2.400
m + 3.184744 g!