52
MPS4232
Section 5: Software

The idencaon sengs control the serial number and the nominal pressure range of the MPS4200.

DESCRIPTION Returns all of the idencaon sengs.
SYNTAX LIST ID
RETURNS All sengs followed by the prompt.
EXAMPLE LIST ID
SET SN 222
SET NPR 5.0000 -5.0000 5.0000 -5.0000
SET MCAST 224.0.1.2
>

COMMAND SN
TYPE Set/Get
DESCRIPTION Sets the serial number of the MPS.
SYNTAX SET SN <sn>
Where:
<sn> is the serial number ranging from 0 to 32767.
RETURNS Prompt when ready to accept a new command
EXAMPLE SET SN 123
>
DEFAULT 100
RANGE 0 to 32767
NOTES This value will get printed to the name of the Cal.cfg le when a SAVE C is issued. Format is:
Cal_<sn>.cfg

COMMAND NPR
TYPE Set/Get
DESCRIPTION Sets the nominal pressure ranges for the MPS
SYNTAX SET NPR <npr1p> <npr1n>
Where:
<npr1p> is the maximum nominal pressure for channels 1 to 32.
<npr1n> is the minimum nominal pressure for channels 1 to 32.
RETURNS Prompt when ready to accept a new command
EXAMPLE SET NPR 15.0 -15.0
>
DEFAULT 15.0 -15.0

COMMAND MCAST
TYPE Set/Get
DESCRIPTION Sets the mulcast IP address used for mulple unit calibraon and validaon. This address must
be the same for all units to be calibrated as a group.