60
MPS4232
Section 5: Software
SYNTAX CALVAL [F]
Where:
F = Iniates a single unit eld calibraon, followed by a eld validaon with no oven connec-
on. MPS connecon to a calibrator required.
When the oponal parameter is not present, the unit performs a full calibraon/validaon.
RETURNS Calibraon and validaon informaon and a prompt when completed.
EXAMPLE CALVAL
[:::data:::]
>
NOTES A Scanivalve SPC4000 or SPC4050 is required for automated calibraons/validaons.
An Ethernet controllable thermal chamber is required for full calibraon/validaons.
COMMAND MCALVAL
TYPE Funcon
DESCRIPTION When the master receives the MCALVAL command, it starts the mulple device full or eld cali-
braon and validaon sequence (connuous), as dened by the variables in LIST C and/or LIST O.
A validaon report is generated and stored in each MPS’s le system.
SYNTAX MCALVAL [F]
F = Iniates a mulple unit single temperature calibraon with no oven connecon. MPS
connecon to a calibrator required.
When the oponal parameter is not present, the unit performs a full mul-unit calibraon and
validaon.
RETURNS Calibraon and validaon informaon and a prompt when completed.
EXAMPLE MCALVAL
[:::data:::]
>
NOTES A Scanivalve SPC4000 or SPC4050 is required for automated calibraons/validaons.
An Ethernet controllable thermal chamber is required for full calibraons/validaons.
COMMAND VAL
TYPE Funcon
DESCRIPTION Starts a full or eld validaon, as dened by the variables in LIST C and/or LIST O. A validaon
report is generated and stored in the MPS’s le system.
SYNTAX VAL [<F or FM>]
Where:
F = Iniates a single temperature eld validaon, with no oven connecon. MPS connecon to
a calibrator required.
FM = Iniates a single temperature manual eld validaon with no oven or calibrator connec-
on.
When the oponal parameter is not present, the unit performs a full validaon.
RETURNS Validaon Data
EXAMPLE VAL
[:::data:::]
>