Section 5: Software
TYPE DISABLED (Set/Get)
DESCRIPTION Sets excitaon mode
SYNTAX SET XITE <polarity>
Where <polarity> is:
0 = Negave only
1 = Posive only
2 = Regular
3 = Reversed
RETURNS Prompt when ready to accept a new command
EXAMPLE SET XITE 2
>
DEFAULT 2
RANGE 0 - 3
WARNING This variable should not be changed from the default seng unless instructed by Scanivalve.

COMMAND ETOL
TYPE Set/Get
DESCRIPTION Changes the error tolerance value set in the validaon report generated by the MPS.
SYNTAX SET ETOL <error>
Where <error> is the value for the Full-Scale error percentage value for pass/fail limits.
When set to ‘0’, the MPS will use the default error percentage based on the pressure range (see
specicaons in Secon 1 for accuracy error percentages).
RETURNS Prompt when ready to accept a new command
EXAMPLE SET ETOL 0.1
>
DEFAULT 0
RANGE 0 - 100