Section 5: Software


COMMAND CAL
TYPE Funcon
DESCRIPTION Starts a full or eld calibraon, single unit or mulple unit, as dened by the variables in LIST C
and/or LIST O.
SYNTAX CAL [<F or FM>]
Where:
F = Iniates a single temperature eld calibraon, with no thermal chamber connecon. MPS
connecon to a calibrator is required.
*FM = Iniates a manual single unit, single temperature eld calibraon with no thermal
chamber or calibrator connecon.
When the oponal parameter is not present, the unit performs a full calibraon.

RETURNS Calibraon data and a prompt when completed.
EXAMPLE CAL
[:::data:::]
>
NOTES A Scanivalve SPC4000 or SPC4050 is required for automated calibraons.
An Ethernet controllable thermal chamber is required for full calibraons.

COMMAND MCAL
TYPE Funcon
DESCRIPTION When the master receives the MCAL command, it starts the mulple device full or eld calibraon
sequence, as dened by the variables in LIST C and/or LIST O.
SYNTAX MCAL [<F or FM>]
Where:
F = Iniates a mulple unit, single temperature led calibraon with no oven connecon. MPS
connecon to a calibrator required.
*FM = Iniates a manual mulple unit single temperature calibraon with no oven connecon.
Does not connect to a calibrator and pressure must be applied manually.
When the oponal parameter is not present, the unit performs a full mul-unit calibraon.

RETURNS Calibraon informaon and a prompt when completed.
EXAMPLE MCAL
[:::data:::]
>
NOTES A Scanivalve SPC4000 or SPC4050 is required for automated calibraons.
An Ethernet controllable thermal chamber is required for full calibraons.

COMMAND CALVAL
TYPE Funcon
DESCRIPTION Starts a full or eld calibraon and validaon sequence (connuous), as dened by the variables
in LIST C and/or LIST O. A validaon report is generated and stored in the MPSs le system.