63
MPS4232
Section 5: Software
COMMAND MIN
TYPE Set/Get
DESCRIPTION Sets the minimum calibraon and validaon temperature and pressure.
SYNTAX SET MIN <tcal> <pcal> <tval> <pval>
Where:
<tcal> = Minimum temperature value for a calibraon.
<pcal> = Minimum pressure value for a calibraon.
<tval> = Minimum temperature value for a validaon.
<pval> = Minimum pressure value for a validaon.
RETURNS Prompt when ready to accept a new command
EXAMPLE SET MIN 0 -5.5 10 -5.0
>
DEFAULT 0 -15 0 -15
RANGE These values are not range checked.
COMMAND MAX
TYPE Set/Get
DESCRIPTION Sets the maximum calibraon and validaon temperature and pressure.
SYNTAX SET MAX <tcal> <pcal> <tval> <pval>
Where:
<tcal> = Maximum temperature value for a calibraon.
<pcal> = Maximum pressure value for a calibraon.
<tval> = Maximum temperature value for a validaon.
<pval> = Maximum pressure value for a validaon.
RETURNS Prompt when ready to accept a new command
EXAMPLE SET MAX 70 5.5 60 5.0
>
DEFAULT 70 15 70 15
RANGE These values are not range checked.
COMMAND DELAY
TYPE Set/Get
DESCRIPTION Sets the delay in minutes and seconds for the oven and calibrator, respecvely.
SYNTAX SET DELAY <iodelay> <odelay> <cdelay> <spcdelay>
Where:
<iodelay> = Inial oven delay in minutes.
<odelay> = Oven delay in minutes. Seng this value to zero enables a stable query to the MPS
(MPS must be +/-0.125°C for 10 minutes, 120 minute meout).
<cdelay> = Calibrator delay in seconds. Seng this value to zero uses the calibrator’s STABLE
command.
<spcdelay> = Special Delay in seconds. This dwell is between the me between the posive
applied points, and the negave applied points.